One patient, one message a minute, a blood pressure every fifteen minutes, then an alarm.
Tuesday 22 September 2026, intensive care at the Leleux Clinic. Jean Lambert has been in bed 4 since the previous evening, record 3072, visit V2026-0921. His monitor continuously measures heart rate, breathing and oxygen saturation. Every 15 minutes, the cuff inflates and takes his blood pressure: the non-invasive blood pressure, or NIBP.
We follow one hour, from 10 to 11 am, message by message. The cast:
CENTRAL: the central monitoring station, which sends one ORU^R01 a minute;
ICUCHART: the ICU charting system, which receives them;
ALERTMGR: the alert manager, which relays alarms to phones.
The central station’s messages follow the IHE device profile, built on v2.6 (MSH-12).
The hour at a glance
10 to 11 am, bed 4
Central station
CENTRAL
Alert manager
ALERTMGR
ICU charting
ICUCHART
10:00 · ORU^R01
10:01 · ORU^R01 + NIBP
10:16 · ORU^R01 + NIBP
10:32:05 · ORU^R40 start
10:32:31 · ORA^R41
10:32:45 · ORU^R40 end
10:33 · ORU^R01
10:59 · ORU^R01
Two feeds live side by side: measurements, to the charting system, every minute; alarms, to the alert manager, only when something happens.
10:01: the first NIBP
The minute message: the device lines, three continuously measured values and the NIBP taken 30 seconds earlier.
ORU^R01^ORU_R01· Observation result (unsolicited)
|||||||||||||||||||
||||||||
|||
|||||||
||||||||||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
||||||||||||||
||||||||||||||
||||||||||||||
Read it top to bottom:
The lines ending in zeros describe the monitor and its modules: ECG, SpO2, NIBP.
OBR-7 gives the default time for every value: 10:01:00.
Heart rate, respiration and SpO2 have no OBX-14: they are continuous measurements, timed by the OBR.
The three NIBP OBX segments do have an OBX-14: 10:00:30. This is an episodic measurement, and its own time overrides the OBR time.
OBX-11 is R: raw values, not yet validated by the nurse.
OBX-7 holds the alarm limits set on the monitor, not lab reference ranges.
10:02: the same NIBP, again
One minute later, the central station resends the last known NIBP. Compare OBX-14 with the previous message: identical.
The next message: the continuous values have moved, the NIBP is the same reading as at 10:01.
ORU^R01^ORU_R01· Observation result (unsolicited)
|||||||||||||||||||
||||||||
|||
|||||||
||||||||||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
||||||||||||||
||||||||||||||
||||||||||||||
The 10:00:30 NIBP therefore travels in 15 messages, until the next one. A naive receiver plots 15 blood pressures. A good receiver deduplicates on the code and OBX-14: one reading, one row. Over the hour: 60 ORU messages, but 4 NIBP readings.
10:32: the alarm
At 10:32:05, heart rate drops to 38. The alarms lesson took apart the start message and the acceptance. Here is the third one, which closes the alarm:
40 seconds after it started, heart rate has recovered: the central station closes alarm ALM-0412.
ORU^R40^ORU_R40
|||||||||||||||||||
||||||||
|||
|||||||||||||||||||||||||||||
||||||||||||||
||||||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
OBR-3 carries a new number, ALM-0413. OBR-29 points back to the original alarm, ALM-0412. The phase is end, the state inactive. The alert manager knows it can stop chasing the team.
What the charting system never saw
And the ICU charting system? It does not receive the alarm. And in the minute feed, it sees nothing: the 10:32:00 snapshot was taken five seconds before the drop, the 10:33:00 one after the recovery (61). Forty seconds of bradycardia fall between two samples. A one-minute feed samples; it does not summarize.
The consequence is clinical. The physician reading the flowsheet at noon finds no value below 50. Nothing tells them Jean Lambert dropped to 38. Only the alarm, the ORU^R40, saw the episode.
Two levers, to decide with the unit and write into the specification:
Subscribe the charting system to alarms. The IHE alarm profile defines an Alert Consumer actor: the charting system can receive the ORU^R40 messages and keep the episode, with its start and end.
Pick the interval knowingly. At 15 seconds, the 10:32:15 and 10:32:30 snapshots would have caught the bradycardia, for four times as many messages. But even at 15 seconds, a 10-second episode can slip between two points. The interval draws a trend; the alarm is what makes sure nothing is missed.
Three traps, and how an integrator spots them
The clock offset. Imagine the central station clock is 7 minutes slow. Nothing breaks: every message is valid. But the chart shows the 10:45:30 NIBP at 10:38:30, and values no longer line up with the nurse’s notes or with other devices that kept the right time. You spot it in the engine log, by comparing MSH-7 with the arrival time:
MSG00513 MSH-7 10:33:00 received 10:40:02MSG00514 MSH-7 10:34:00 received 10:41:01
Network delay varies. A fixed 7-minute gap is a clock. The fix: NTP everywhere, and timestamps with a time zone offset.
The patient nobody linked to the monitor. A second patient arrives in bed 5. Here, association is manual. The nurse plugs in the monitor but has not yet associated the patient at the central station. Messages go out anyway:
The bed 5 monitor sends its values, but no patient is associated with it at the central station.
ORU^R01^ORU_R01· Observation result (unsolicited)
|||||||||||||||||||
|||
|||
|||||||
||||||||||||||||||
|||||||||||
|||||||||||
|||||||||||
PID-3 is empty. The charting system cannot attach these values to anyone. Depending on the central station, either nothing is sent, or data goes out “to the bed”. You spot it through a growing orphan queue, or a gap in the new patient’s chart. The fix: bed association is part of the admission routine, and the receiver never guesses the patient.
Demo mode. A technician swaps the bed 4 monitor for a loaner unit that was left in demo mode. The patient is correctly associated. The values are perfect and flat:
The loaner unit sends demonstration values under Jean Lambert’s identity.
ORU^R01^ORU_R01· Observation result (unsolicited)
|||||||||||||||||||
||||||||
|||
|||||||
||||||||||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||
Two signals: MSH-11 = D and DEMO in OBX-8. A good feed sets them. A poor one sets nothing, and all you have left is your eye: the same value, minute after minute, when a real patient always varies a little. The fix: the charting system refuses anything that is not P or that carries DEMO or TEST, and biomedical engineering checks the mode before installing a loaner.
The hour in numbers
What
How many
ORU^R01 from the central station
60, one per minute
Distinct NIBP readings
4, each repeated until the next
Alarms
1: one ORU^R40 start, one ORA^R41, one ORU^R40 end
ACK messages
one per message received
Check your understanding
1 of 5
Between 10 and 11 am, how many distinct non-invasive blood pressure (NIBP) readings should the charting system show?