Module 8 · Patient monitoring·11 min
Alarms
A bradycardia, an alarm message, an acknowledgement: what happens in a few seconds.
10:32 in the intensive care unit at the Leleux Clinic. The heart rate of Jean Lambert, bed 4, drops to 38 per minute. The monitor sounds at the bedside. In under a second, the central station sends a message saying: “alarm, bed 4, low heart rate”.
Here is that message. Click OBX-5 on the third OBX line: start.
The central station tells the alert manager that a low heart rate alarm has just started at bed 4.
One message, one alarm
MSH-9 is ORU^R40: an unsolicited alert observation. It entered the standard in v2.8, together with its response ORA^R41. The IHE profile that uses it, ACM (Alert Communication Management), is built on v2.6 and adopted it early. Hence 2.6 in MSH-12.
The golden rule of this profile: one alarm per message. An ORU^R40 never mixes two alarms, and it does not carry the measurement feed. It lives next to the minute-by-minute ORU^R01 feed without replacing it.
The structure is still a result. PID and PV1 say who and where: patient 3072, bed ICU^04^A^LELEUX. The OBR announces an alarm (196616^MDC_EVT_ALARM^MDC) and gives it a unique number in OBR-3: ALM-0412. Then each OBX describes one facet of the alarm.
The facets, OBX by OBX
OBX-4 ends with a facet number (.1 to .7). Everything before it, 1.1.1.147842, says where the alarm comes from: the device, the module, the channel (see the lesson on reading a monitoring ORU), then the MDC code of the measurement involved, 147842 for heart rate. It is no longer a simple rank as in the ORU^R01. The facet number is kept mainly for backward compatibility: the code in OBX-3 is what identifies the facet.
| Facet | OBX-3 |
Here | What it says |
|---|---|---|---|
| 1 | MDC_EVT_LO |
Low HR | which event: a value below its low limit |
| 2 | MDC_ECG_HEART_RATE |
38, limits 50-120 | which measurement, its value, its alarm limits in OBX-7 |
| 3 | MDC_ATTR_EVENT_PHASE |
start | where the alarm stands |
| 4 | MDC_ATTR_ALARM_STATE |
active | the state at the monitor |
| 5 | MDC_ATTR_ALARM_INACTIVATION_STATE |
enabled | is the sound silenced at the bedside? |
| 6 | MDC_ATTR_ALARM_PRIORITY |
PH | the priority |
| 7 | MDC_ATTR_ALERT_TYPE |
SP | physiological, technical or advisory |
The numeric codes (196670, 147842) are MDC codes, the device nomenclature (IEEE 11073). The text “Low HR” is only a label, translated into whatever language the central station runs in. An integrator always maps the code, never the label.
Three short lists are enough to read almost any alarm:
- Priority:
PHhigh,PMmedium,PLlow,PNnot indicated. - Type:
SPphysiological alarm (the patient),STtechnical alarm (lead off, battery),SAadvisory. - Phase:
start(an end is expected),start_only(no end expected),continue,end,present, plus a few others for escalation or silencing.
Our example is a limit alarm: heart rate drops below the set threshold. In cardiac monitoring, the alarms that matter most are often arrhythmia alarms: asystole, ventricular fibrillation or tachycardia. They have their own MDC event codes in facet 1, for example 199684 MDC_EVT_ECG_ASYSTOLE. They are not limit violations, but the message structure stays the same.
From the monitor to the nurse’s phone
The alarm does not stop at the central station. ACM defines three roles: the reporter (here the central station), the alert manager, which decides who to notify, and the communicator, which pushes the notification to a phone or pager. Play the animation:
Monitor
MONITOR, bed 4
Central station
CENTRAL
Alert manager
ALERTMGR
Phone
nurse
- HR 38/min
- ORU^R40 · start
- ACK · CA
- Notification
- Accepted
- ORA^R41 · ACCEPTED
- ORU^R40 · end
Notice the two levels of response. The ACK with CA comes back at once: it is the accept acknowledgement of enhanced mode (see the lesson on acknowledgements). The real business answer, “someone is on it”, can take 20 seconds or 3 minutes. So it travels in a separate message, the ORA^R41. To receive it, the central station set AL in MSH-16.
The acceptance message
The alert manager tells the central station that the nurse has accepted the bed 4 alarm.
Three fields make the link:
MSA-2copies theMSH-10of the message that opened the alarm,MSG00610. Never the ID of an update.OBR-29(parent) carries the alarm numberALM-0412. Same value as the originalOBR-3, but written as subcomponents, with&.- The
PRTsegment says who and what:AAP(the person who took the alarm), Julie Peeters, responseACCEPTED, at 10:32:31. This segment comes from v2.7; the profile borrows it.
One ORU^R40 can trigger several ORA^R41: “received”, “delivered”, “read”, “accepted”. And if nobody answers, the manager can escalate to someone else.
The end of the alarm
At 10:32:45, the heart rate is back to 58. The central station sends one last ORU^R40. Same structure, three differences: a new number in OBR-3 (ALM-0413), the link to the original alarm in OBR-29, and phase end with state inactive. The receiver ties it all together through ALM-0412.
And in FHIR?
The value that triggered the alarm becomes an Observation: LOINC code 8867-4 (heart rate), MDC code 147842 as a second coding, UCUM unit /min, and a “low” interpretation. Modelling the alarm itself, with its phases and responses, is beyond this lesson.
Check your understanding
1 of 4You receive an ORU^R40 where the OBX 68481^MDC_ATTR_EVENT_PHASE^MDC has OBX-5 = end. What does this message tell you?