The Path
Module 8 · Patient monitoring·12 min
Reading a monitoring ORU
Several devices, a hierarchy in OBX-4, the time of each measurement in OBX-14.
An intensive care patient is never connected to just one device. The monitor measures the heart, oxygen and blood pressure. A pump infuses on two channels. It all arrives in the same ORU, and the EHR must know where each value comes from, and when it was measured.
At 9:31, a two-channel syringe pump is connected to Jean Lambert. Here the central station also acts as the device gateway: it groups the monitor and the pump in the same message, in full form, with the device hierarchy. In real sites, pumps often report through their own server or a device integration platform. The message works the same way.
One ORU for two devices at bed 4. Click the OBX-4 values one after the other: they draw the device tree.
Seventeen OBX, but only six values. The other eleven measure nothing: they describe the devices.
One OBR, one moment
Start with OBR-7: 9:31. It is the default time for everything that follows. An OBX without a time of its own is a 9:31 measurement. Keep that in mind, we will come back to it with blood pressure.
OBX-4: the device tree
The IHE profile uses OBX-4 in a precise way: four numbers separated by dots.
| Level | Meaning | In our message |
|---|---|---|
| MDS (Medical Device System) | the whole device | 1 = the monitor, 2 = the pump |
| VMD (Virtual Medical Device) | a sub-device, a module | 1.1 = ECG, 1.2 = SpO2, 1.3 = non-invasive pressure |
| CHAN | a channel | 2.1.1 = channel 1, 2.1.2 = channel 2 |
| METRIC | the measurement itself | 1.1.1.1 = heart rate |
Two kinds of lines are mixed. A device line ends in zeros (1.0.0.0, 1.3.0.0): it announces a level of the tree, with no value, and X in OBX-11. A measurement line has all four levels filled and a value in OBX-5.
Read the message like a table of contents: monitor, then ECG, then ECG channel, then heart rate. Then the oximeter and its SpO2. Then pressure. Then the pump, channel 1, channel 2.
The pump twins
Look at OBX 15 and 17. Same code in OBX-3, 157784, the pump flow rate. Two values: 5 and 12 mL/h. Without OBX-4, there is no way to know which channel runs at which rate. With OBX-4, it is obvious: 2.1.1.1 is channel 1, 2.1.2.1 channel 2.
The previous lesson showed the flattened form: every measurement hangs directly off the monitor, as 1.0.0.1, 1.0.0.2. That is enough for one simple device. The IHE profile lists its limits: it cannot tell pump channels apart, and two measurements with the same code risk being lost.
Careful: the numbers are not normative. 1.1.1.1 is not “heart rate” forever. The profile states that they may change from one message to the next. You read the code in OBX-3; OBX-4 only says where it sits.
OBX-14: the time of each measurement
Now, blood pressure. OBX 10 and 11 carry 20260922091530+0200 in OBX-14. Non-invasive pressure is taken when the cuff inflates, for example every fifteen minutes. It is not a continuous measurement: it is an episodic one.
| Kind | Example | Time of the value |
|---|---|---|
| Periodic | heart rate, SpO2, flow rate | OBR-7, the time of the message |
| Episodic | cuff pressure | OBX-14, which overrides OBR-7 |
The IHE profile puts it this way: an OBX-14 signals an episodic measurement, its absence a periodic one.
Monitor
cuff
Central station
one ORU a minute
EHR
deduplicates
- NIBP 128/74
- ORU · OBX-14 9:15:30
- ORU · OBX-14 9:15:30
- NIBP 131/76
- ORU · OBX-14 9:45:30
If the central station resends the last NIBP in every message, the receiver has to deduplicate: same code in OBX-3, same time in OBX-14, same measurement. Try it below.
The 9:15 NIBP is sent again. Change the OBX-14 time to 20260922093200+0200: the EHR would believe in a new reading at 9:32 that never happened.
And in FHIR?
Each measurement line becomes an Observation. OBX-14, or else the OBR time, gives effectiveDateTime. OBX-18 gives Observation.device, a Device resource found by the device identifier. For the hierarchy, the official v2-to-FHIR guide leaves OBX-4 to the project. The FHIR Point-of-Care Device guide rebuilds the tree, from the measurement up to the whole device.
Two exceptions worth knowing:
- Blood pressure. The FHIR blood pressure profile groups systolic and diastolic into one Observation, LOINC code
85354-9, with two components:8480-6and8462-4. So OBX 10 and 11 become one Observation, not two. - OBX-7. The v2-to-FHIR guide sends it to
referenceRange. In monitoring, OBX-7 holds the alarm limits set on the monitor, not a normal range. Do not copy them intoreferenceRange: keep them out of the Observation, or in a documented extension.
Check your understanding
1 of 5An OBX has OBX-4 = 1.3.0.0 and nothing in OBX-5. What does it describe?
+50 XPPass the quiz (4 correct answers) to complete the lesson.
Reviewed and checked on 28 September 2026