Module 8 · Patient monitoring·10 min

The vital signs feed

A central station sending results nobody asked for, every minute.

In intensive care, nobody orders “a heart rate at 9:31”. The patient is connected, the monitor measures continuously, and the EHR receives the values every minute. No request, no planned end: a feed.

In the sleep lab, Marie Dubois’s SpO2 was recorded all night, then summed up in a single ORU. In intensive care, the same SpO2 goes out every minute, for days.

Jean Lambert, 68, has been in bed 4 of the Leleux Clinic intensive care unit since the day before, for heart rhythm monitoring. Here is the message the central monitoring station sends to the EHR at 9:30.

The central station sends Jean Lambert’s heart rate, SpO2 and respiratory rate from bed 4 to the EHR. Click OBR-4: that is the “order” behind the whole feed.

ORU^R01^ORU_R01· Observation result (unsolicited)
|||||||||||||||||||
||||||||
|||
|||||||
|||||||||||
|||||||||||
|||||||||||
|||||||||||

The six-digit codes in OBX-3 are MDC codes, the language of devices. They get their own lesson later in this module. Here, we look at the feed. Note that ICU^04^A in PV1-3 reads unit, room, bed: in intensive care one room holds one bed, so everyone says “bed 4”.

Nobody asked for anything

In the classic path, a physician orders (ORM), the lab performs the test, then answers (ORU). Monitoring has no order per value. Connecting the patient to the monitor acts as a standing order. The IHE profile that governs this feed says so plainly: the ORU^R01 is sent with or without a prior order.

The message still looks like a result. OBR-4 names a generic service, here the SNOMED CT concept 182777000, “Monitoring of patient”. OBR-7 gives the time of the snapshot: 9:30. Every OBX that says nothing else is a 9:30 measurement.

Who talks to whom

The bedside monitor almost never speaks HL7 itself. It sends its measurements to a central monitoring station over the manufacturer’s network. The central station, or a gateway, translates to HL7 and pushes to the EHR.

System Role Name in the IHE profile
Bedside monitor measures, displays, alarms the device, outside HL7
Central station or gateway groups the beds, writes the ORUs Device Observation Reporter
EHR or ICU charting system receives, displays, archives Device Observation Consumer

The profile is called DEC, Device Enterprise Communication. Its core transaction, PCD-01, is exactly this message: an ORU^R01 carrying measurements. The profile comes from the IHE Devices domain, long called PCD (Patient Care Device): hence IHE PCD in MSH-21.

The reporter is not limited to the monitor. The profile expects it to gather data from every device connected to a patient: monitor, ventilator, pumps. You will see a monitor and a pump in the same message in the next lesson.

Three minutes at bed 4

Monitor

bed 4

Central station

writes the HL7

EHR

receives

  1. measurements
  2. ORU^R01 · 9:30
  3. ACK · CA
  4. ORU^R01 · 9:31
  5. ACK · CA
  6. ORU^R01 · 9:32

Every minute, and nothing made up

The IHE profile allows a wide range: several messages per minute for the most unstable patients, down to one every 24 hours for home follow-up. The typical interval is one minute. It is set at installation.

Each message is a snapshot. If a sensor falls off, the value is missing: the central station must not interpolate. The profile forbids it in so many words. A gap in the trend tells the caregiver something; a made-up value lies to them.

What that means in volume

One value a minute sounds like little. Let’s do the math for a 30-bed unit, one message per bed per minute.

Period 1 bed 30 beds
1 hour 60 messages 1,800 messages
1 day 1,440 messages 43,200 messages
1 year (365 days) 525,600 messages 15,768,000 messages

With about twenty OBX per message, the day’s 43,200 messages hold 864,000 OBX. Move the interval to 15 seconds and everything is multiplied by 4: 172,800 messages a day. That is why the interval is chosen for what the EHR needs, not for what the central station can do.

In our examples, the bed 4 feed has its own numbering: one MSH-10 a minute, MSG00300 at 7 am, MSG00450 at 9:30. Alarms, which go to another system, have their own sequence.

Acknowledgements: just “got it”

Look at the end of MSH: AL in MSH-15, NE in MSH-16. The IHE profile sets both values for PCD-01. The EHR therefore returns a technical acknowledgement, MSA-1 = CA, as soon as it has taken the message in. No application acknowledgement is expected. At 1,800 messages an hour, a slow receiver quickly builds a queue.

Another version: 2.6

MSH-12 is 2.6, and MSH-21 carries the IHE profile identifier (1.3.6.1.4.1.19376.1.6.1.1.1). The DEC profile is built on HL7 v2.6, so our monitoring messages follow it. For the fields in this module, you read them the same way as in 2.5.1, which remains the reference for the rest of the course. One exception: the alarms’ PRT segment, which only exists from 2.7 on.

What this feed does not carry

The waveform. The continuous ECG the nurse watches at the central station stays on the manufacturer’s network: the DEC profile does not cover waveforms. Only waveform snippets can travel with an alarm, in the alarm profile (later lesson). Alarms themselves, arrhythmia alarms included, also travel separately. The PCD-01 feed carries values only: rates, saturation, pressures, flow rates.

And in FHIR?

Each measurement OBX becomes an Observation in the vital-signs category (blood pressure is the exception, see the next lesson). The FHIR vital signs profile requires a LOINC code (8867-4 for heart rate, unit /min). The device’s MDC code can follow as an additional code. One catch: the official v2-to-FHIR guide gives no equivalent for status R. The project has to choose, here preliminary.

{
  "resourceType": "Observation",
  "status": "preliminary",
  "category": [{ "coding": [{ "system": "http://terminology.hl7.org/CodeSystem/observation-category", "code": "vital-signs" }] }],
  "code": { "coding": [
    { "system": "http://loinc.org", "code": "8867-4", "display": "Heart rate" },
    { "system": "urn:iso:std:iso:11073:10101", "code": "147842", "display": "MDC_ECG_HEART_RATE" }
  ] },
  "subject": { "reference": "Patient/3072" },
  "effectiveDateTime": "2026-09-22T09:30:00+02:00",
  "valueQuantity": { "value": 72, "unit": "/min", "system": "http://unitsofmeasure.org", "code": "/min" }
}

Check your understanding

1 of 4

No order was ever sent for bed 4. Yet an ORU^R01 arrives every minute. Is that normal?

+50 XPPass the quiz (3 correct answers) to complete the lesson.

Reviewed and checked on 28 September 2026